| Model Number |
ZA9003 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Age, weight, ethnicity: per regulation eu (b)(4) (general data protection regulation), patient identifiers were not collected or recorded and therefore are not available.Date of event: date unknown/ not provided.Implant date: unknown/ not provided.Explant date: unknown/ not provided.Initial reporter telephone number: (b)(6).The device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of the device history record, complaint trending, and risk documentation for this device will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.Attempts have been made to obtain missing information.However, no further information was available.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during implantation/application the intraocular lens (iol) haptic (leg) was broken.Product availability is unknown and no further details are available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: section d9 - device available for evaluation? yes.Section d9 - date returned to manufacturer:3rd august2022.Section h3 - device evaluated by manufacturer? yes.Device evaluation: visual inspection under magnification revealed that the complaint lens was received coated in viscoelastic residue and with a detached haptic.The remaining portion of the detached haptic was bent.The lens was cleaned and, it was revealed to have been cut.The complaint issue was identified during product evaluation; however, based on the complaint investigation results the complaint issue could not be confirmed to be related to the manufacturing or design process.The other observed issues during the product evaluation could not be confirmed to be related to a manufacturing or design issue.Therefore, there is no indication of a product deficiency or product malfunction.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|