• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CHOICE; WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CHOICE; WIRE, GUIDE, CATHETER Back to Search Results
Model Number 1960
Medical Device Problem Codes Failure to Advance (2524); Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/17/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that coating issue occurred.The 80% stenosed target lesion was located in the non-tortuous and non-calcified left anterior descending artery.A 182cm (5pk) choice guidewire was selected for use.However, resistance was encountered when the device was tried to cross the lesion and the tip of the wire was peeled off.The procedure was completed with another wire of same brand.There were no patient complications reported.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: unit returned with its original pouch overall visual revision did not identify failures or evidence that could be lost due to decontamination process.Spring tip was found stretched and kinked.No abnormalities were noticed int the coating of the guidewire.Boston scientific manufacturing processes include extensive inspections to ensure that all finished devices meet specifications.
 
Event or Problem Description
It was reported that coating issue occurred.The 80% stenosed target lesion was located in the non-tortuous and non-calcified left anterior descending artery.A 182cm (5pk) choice guidewire was selected for use.However, resistance was encountered when the device was tried to cross the lesion and the tip of the wire was peeled off.The procedure was completed with another wire of same brand.There were no patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHOICE
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15094893
Report Number2124215-2022-25446
Device Sequence Number14761086
Product Code DQX
UDI-Device Identifier08714729150558
UDI-Public08714729150558
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K143587
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/14/2023
Device Model Number1960
Device Catalogue Number1960
Device Lot Number0028559214
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/07/2022
Supplement Date Received by Manufacturer11/02/2022
Initial Report FDA Received Date07/25/2022
Supplement Report FDA Received Date11/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/14/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
Patient RaceAsian
-
-