STRYKER NEUROVASCULAR CORK TARGET 360 NANO 1.5MM X 3CM; DEVICE, NEUROVASCULAR EMBOLIZATION Back to Search Results
Model Number M0035421530
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/28/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Discarded by hospital.
 
Event or Problem Description
It was reported during the coil embolization procedure to treat bleeding due to dissection of spinal muscular atrophy (sma) branch, the subject coil was inserted into the microcatheter.The delivery wire of the subject coil was kinked and detached when it was inserted into the coil detachment device.The broken delivery wire was removed from the detachment device and subject coil was removed from the patient¿s anatomy.The physician replaced it with a new device and continued the procedure without clinical consequences to the patient.
 
Additional Manufacturer Narrative
H4 manufacturing date ¿ added, d4 expiration date - added.Due to the automated manufacturing execution system (mes) there are controls in the manufacturing process to ensure the product met specifications upon release.The subject device is not available; therefore, visual and functional testing as well as physical analysis cannot be performed.The reported complaint could not be confirmed, and it could not be definitively determined if the device failed to meet specifications because the product was not returned.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.Additional information provided by the customer indicated that continuous flush set up and maintained throughout the procedure.No damage was noted to the packaging prior to opening, the device was confirmed to be in good condition during preparation/prior to use on the patient and was prepared as per the dfu.The delivery wire was kinked and detached when it was inserted into the inzone.The broken delivery wire was removed from inzone, and the coil was removed.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause and the device was not returned, an assignable cause of undeterminable will be assigned to this complaint h3 other text : device not returned for evaluation.
 
Event or Problem Description
It was reported during the coil embolization procedure to treat bleeding due to dissection of spinal muscular atrophy (sma) branch, the subject coil was inserted into the microcatheter.The delivery wire of the subject coil was kinked and detached when it was inserted into the coil detachment device.The broken delivery wire was removed from the detachment device and subject coil was removed from the patient¿s anatomy.The physician replaced it with a new device and continued the procedure without clinical consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET 360 NANO 1.5MM X 3CM
Common Device Name
DEVICE, NEUROVASCULAR EMBOLIZATION
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI  NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI   NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key15095056
Report Number3008881809-2022-00368
Device Sequence Number5207924
Product Code HCG
UDI-Device Identifier04546540697943
UDI-Public04546540697943
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K153658
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/28/2024
Device Model NumberM0035421530
Device Catalogue NumberM0035421530
Device Lot Number23178609
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/29/2022
Supplement Date Received by Manufacturer09/12/2022
Initial Report FDA Received Date07/25/2022
Supplement Report FDA Received Date09/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured09/30/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
INZONE POWER SUPPLY (STRYKER)
Patient SexUnknown
-
-