ABBOTT VASCULAR ARMADA 18 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number 1013470-060
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/05/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is not returning for analysis.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.The armada 14 referenced is filed under a separate medwatch report number.
 
Event or Problem Description
It was reported that the procedure was to treat a proximal superficial femoral artery (sfa) to distal popliteal.The 3.0x60mm armada 18 and 4.0x60mm armada 14 balloon dilatation catheter both ruptured.The armada 14 ruptured at 11 atmospheres; however, both balloons were noted to rupture before reaching 14 atmospheres.Another unspecified balloon was used to successfully complete the procedure.There was no adverse patient effect reported and no clinically significant delay reported.No additional information was provided.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported balloon rupture.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 18 PTA CATHETER
Common Device Name
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005718570 (P099)
cashel road
clonmel tipperary
EI  
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key15095275
Report Number2024168-2022-08106
Device Sequence Number17442534
Product Code LIT
UDI-Device Identifier08717648208409
UDI-Public08717648208409
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K151317
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2023
Device Model Number1013470-060
Device Catalogue Number1013470-060
Device Lot Number0042841
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/05/2022
Supplement Date Received by Manufacturer07/26/2022
Initial Report FDA Received Date07/25/2022
Supplement Report FDA Received Date08/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/28/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-