| Model Number |
453564667151 |
| Medical Device Problem Codes |
Measurement System Incompatibility (2982); Application Program Freezes or Fails to Launch (4031)
|
| Health Effect - Clinical Codes |
Shock (2072); Insufficient Information (4580)
|
| Date of Event |
06/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A follow-up report will be submitted upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
The customer reported monitoring disruptions during defibrillation.We are in the process of obtaining additional information from the customer regarding patient involvement.
|
| |
|
Event or Problem Description
|
|
The customer reported monitoring disruptions during cautery use and defibrillation.During an angiogram of the coronaries with angioplasty and stenting, the patient went into ventricular fibrillation (v-fib) and had to be defibrillated twice.The flex cardio device was in-use during the time of the incident during the first defibrillated shock, the flex cardio stopped working.The customer indicated they rebooted the flex cardio to resolve the issue.A philips field service engineer (fse) went to the customer site and was unable to confirm the flex cardio failure as reported, therefore we are considering this to be a malfunction of cause unknown.The device passed all functional testing and was returned to service.
|
| |
|