Model Number 1987-06-010
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Unspecified Musculoskeletal problem (4535)
Date of Event 06/13/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Implantation of an lps knee prosthesis: xsm femur and femor lps + 10 mm adapter (ref without intraoperative complication.Detection in consultation of a rotation disorder of the femur, which was not present on the intraoperative pictures related to a mismatch of the morse cone (either between the sleeve and the adapter or between the adapter and the femur).Major rotation disorder in external rotation of the femur with severe alteration of the flexion axis of the leg.Surgical revision for replacement of adapter and femoral implant (same manufacturer, same sizes).No intraoperative complications.
 
Event or Problem Description
Additional information received.A.What is the affected side: right knee.B.Was there a surgical delay? if yes, what was the duration of the delay? no delay, during the initial surgery.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary: the device associated with this complaint was returned for examination.Visual examination found, the device still firmly attached to the femoral component.And nothing indicative of a device nonconformance could be seen.Not confirming, the reported allegation.However, there are circular marks around the device.Suggesting, it was attached and detached from a sleeve mating component.No device associated with this report was received for examination.Depuy synthes considers, the investigation closed at this time.Should additional information be received, the information will be reviewed.And the investigation may be re-opened as necessary.Device history lot: a search of the depuy nonconformance (nc) quality system, found no nc¿s associated with this product/lot combination.Based on the inability to find any nc¿s against the provided product code/lot code combination, it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LPS FEM TO SLEEVE ADAPTER +10
Common Device Name
LPS AND S-ROM : KNEE FEMORAL ACCESSORY
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key15099306
Report Number1818910-2022-13954
Device Sequence Number4149454
Product Code MBH
UDI-Device Identifier10603295078500
UDI-Public10603295078500
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K071417
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1987-06-010
Device Catalogue Number198706010
Device Lot NumberJ6414R
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/07/2022
Supplement Date Received by Manufacturer11/04/2022
Initial Report FDA Received Date07/25/2022
Supplement Report FDA Received Date11/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/02/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
LPS DISTAL FEM COMP XSM RT (EX PKT); LPS FEM TO SLEEVE ADAPTER +10; LPS UNIV TIB HIN INS XSM 14MM; SROM*NRH REPL HINGE PIN/XSM/SM; UNIVERSAL FEM SLV DIS POR 31MM
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-