• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US LPS DISTAL FEM COMP XSM RT; LPS AND S-ROM : KNEE FEMORAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US LPS DISTAL FEM COMP XSM RT; LPS AND S-ROM : KNEE FEMORAL Back to Search Results
Model Number 1987-14-105
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Unspecified Musculoskeletal problem (4535)
Date of Event 06/13/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Implantation of an lps knee prosthesis: xsm femur and femor lps + 10 mm adapter (ref without intraoperative complication.Detection in consultation of a rotation disorder of the femur, which was not present on the intraoperative pictures related to a mismatch of the morse cone (either between the sleeve and the adapter or between the adapter and the femur).Major rotation disorder in external rotation of the femur with severe alteration of the flexion axis of the leg.Surgical revision for replacement of adapter and femoral implant (same manufacturer, same sizes).No intraoperative complications.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary the device associated with this complaint was received for examination.Visual examination found nothing indicative of a device nonconformance, the reported allegation cannot be confirmed.Some slight extraction markings can be seen near the sleeve adapter mating component.No device associated with this report was received for examination.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot a search of the depuy nonconformance (nc) quality system found no nc¿s associated with this product/lot combination.Based on the inability to find any nc¿s against the provided product code/lot code combination, it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.
 
Event or Problem Description
Additional information received.A.What is the affected side: -right knee.B.Was there a surgical delay? if yes, what was the duration of the delay? -no delay during the initial surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LPS DISTAL FEM COMP XSM RT
Common Device Name
LPS AND S-ROM : KNEE FEMORAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key15099353
Report Number1818910-2022-13955
Device Sequence Number19887089
Product Code JWH
UDI-Device Identifier10603295078708
UDI-Public10603295078708
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K063686
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1987-14-105
Device Catalogue Number198714105
Device Lot NumberJP9459
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/07/2022
Supplement Date Received by Manufacturer11/04/2022
Initial Report FDA Received Date07/25/2022
Supplement Report FDA Received Date11/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/06/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
LPS DISTAL FEM COMP XSM RT (EX PKT); LPS FEM TO SLEEVE ADAPTER +10; LPS UNIV TIB HIN INS XSM 14MM; SROM*NRH REPL HINGE PIN/XSM/SM; UNIVERSAL FEM SLV DIS POR 31MM; UNKNOWN KNEE FEMORAL SLEEVE
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-