• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL AMPLATZER SEPTAL OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL AMPLATZER SEPTAL OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER Back to Search Results
Model Number SEPTAL OCCLUDER
Medical Device Problem Code Migration or Expulsion of Device (1395)
Health Effect - Clinical Codes Mitral Perforation (2512); Cardiac Perforation (2513); Pericardial Effusion (3271)
Date of Event 06/30/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that on (b)(6) 2022, a 19mm amplatzer septal occluder (aso) was implanted using a 9f amplatzer trevisio intravascular delivery system.The aso had been sized using a 24mm amplatzer sizing balloon and ice examination it was confirmed via intracardiac echocardiography (ice) that 15 seconds after release the aso embolized into the left atrium.In addition, the patient was found to have pericardial effusion.The physician attempted to retrieve the device via snare, however was unsuccessfully.The patient then underwent emergent open heart surgery to have the atrial septal defect repaired, aso removed, pericardial effusion drained, and perforation repaired.It was noted during the explant procedure that at an unknown point, the patient's mitral valve had sustained an injury.The patient was sent to another hospital and put on extracorporeal membrane oxygenation (ecmo); the patient is waiting to undergo surgery to treat the injured mitral valve.
 
Additional Manufacturer Narrative
An event of embolism of the device, pericardial effusion, perforation, and damage to the mitral valve was reported.Information from the field indicated that the device migrated 15 minutes after it was implanted, and that the device was not thought to be the cause of the pericardial effusion or perforation.The physician though the cause of the device embolism could be insufficient rims or the device being too small.A returned device assessment, to measure the device, could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported event could not be conclusively determined.There is no indication of a product quality issue with regards to manufacture, design, or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Common Device Name
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL #3014918977
edificio #44 calle 0, ave. 2
el coyol alajuela 1897- 4050
CS   1897-4050
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key15099426
Report Number2135147-2022-00568
Device Sequence Number5174731
Product Code MLV
UDI-Device Identifier00811806010168
UDI-Public00811806010168
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P000039
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/30/2024
Device Model NumberSEPTAL OCCLUDER
Device Catalogue Number9-ASD-019
Device Lot Number7310866
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/30/2022
Supplement Date Received by Manufacturer08/31/2022
Initial Report FDA Received Date07/25/2022
Supplement Report FDA Received Date09/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/12/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AMPLATZER TREVISIO INTRAVASCULAR DS,9-ATV09F45/80
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexFemale
-
-