Product complaint #: (b)(4).Device not returned.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent.Procedure name and date? event date? device return status?.
|
|
Product complaint # (b)(4).Additional information has been requested and received.If further details are received at a later date a supplemental medwatch will be sent.Procedure name and date? - unk event date? - unk device return status? the sample has been sent.H3 evaluation: the product was returned for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that one needle/suture piece of product code was received for evaluation.The product code is double-armed.In order to evaluate the conditions of the returned sample, on the needle-suture piece, the end was cut; after further evaluation, crimping marks were observed near the end suture possibly caused by a surgical instrument.As with any device, care should be taken to avoid damage to the strand when handling.Avoid the crushing or crimping action of the surgical instruments, such as needle holders and forceps.The condition of the sample received indicates improper handling of the device.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|