• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION EXTRACTOR PRO RX; CATHETER, BILIARY, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION EXTRACTOR PRO RX; CATHETER, BILIARY, DIAGNOSTIC Back to Search Results
Model Number M00547000
Medical Device Problem Code Deflation Problem (1149)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/29/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device has been received for analysis; however, the analysis has not yet been completed.Upon completion of the device analysis, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that an extractor pro rx retrieval balloon was used in the common bile duct (cbd) during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2022.During the procedure, the balloon was able to be inflated; however, it could not be deflated.The device was removed with the scope from the patient, and the device was cut in order to be removed from the scope.The procedure was completed with another extractor pro rx retrieval balloon.There were no patient complications as a result of this event.
 
Event or Problem Description
It was reported to boston scientific corporation that an extractor pro rx retrieval balloon was used in the common bile duct (cbd) during an endoscopic retrograde cholangiopancreatography (ercp) procedure performed on (b)(6) 2022.During the procedure, the balloon was able to be inflated; however, it could not be deflated.The device was removed with the scope from the patient, and the device was cut in order to be removed from the scope.The procedure was completed with another extractor pro rx retrieval balloon.There were no patient complications as a result of this event.
 
Additional Manufacturer Narrative
Device code a1401 captures the reportable event of balloon failure to deflate.Investigation results the returned extractor pro rx retrieval balloon was analyzed and found that the balloon and the catheter of device had no damages.Functional analysis could not be performed due to the balloon lumen being occluded by dry contrast, and the occlusion could not be unblocked; this occlusion is normal due to the use of contrast during the procedure.With all available information, boston scientific concludes the reported event of balloon could not be deflated was not confirmed.The balloon was found blocked with dry contrast medium, since the balloon could not be tested it could not be functionally/visually verified.Therefore, the most probable root cause is cause not established.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXTRACTOR PRO RX
Common Device Name
CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
cork business technology park
model farm road
cork
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15099671
Report Number3005099803-2022-03846
Device Sequence Number14739633
Product Code FGE
UDI-Device Identifier08714729790266
UDI-Public08714729790266
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
K102082
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/19/2024
Device Model NumberM00547000
Device Catalogue Number4700
Device Lot Number0029251739
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/18/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/30/2022
Supplement Date Received by Manufacturer08/01/2022
Initial Report FDA Received Date07/25/2022
Supplement Report FDA Received Date08/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/20/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-