DEPUY ORTHOPAEDICS INC US PFC SIGMA MOD TIB TRAY CEM SZ2; SIGMA KNEE PRIMARY : KNEE TIBIAL TRAY Back to Search Results
Model Number 86-6022
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Codes Fall (1848); Pain (1994); Joint Laxity (4526)
Date of Event 07/07/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # pc-(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that pt had done well with primary total knee until experiencing a fall and then became painful.Surgeon noted mild laxity in extension but more significant instability in flexion.A poly exchange was attempted los of extension was noted and patella was riding too low as the knee was flexed when poly thickness was increased to stabilize flexion.All components were removed and a revision total knee was performed.Tibia component was found to be loose when surgeon began to remove.Ultimately, a hinged prosthesis had to be utilized to stabilize the flexion gap and address the joint line for patellar position.There was no patient harm.Instability was not due to wear, malalignment, nor under sizing or over resection.No surgical delay.Doi: (b)(6) 2015.Dor: (b)(6) 2022.Affected side: right.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.All available x-rays were reviewed.Complaint cannot be confirmed, no loose condition was observed on the provided evidence.No evidence anything indicative of a device nonconformance was found.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed, and the investigation may be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PFC SIGMA MOD TIB TRAY CEM SZ2
Common Device Name
SIGMA KNEE PRIMARY : KNEE TIBIAL TRAY
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JJM (SUZHOU) LTD. 3006356043
no.299 changyang st
suzhou industrial park
suzhou jiangsu 21512 -6
CH   21512-6
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key15100025
Report Number1818910-2022-13957
Device Sequence Number17502253
Product Code JWH
UDI-Device Identifier10603295216773
UDI-Public10603295216773
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K884796
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number86-6022
Device Catalogue Number866022
Device Lot NumberD15051746
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/07/2022
Supplement Date Received by Manufacturer09/02/2022
Initial Report FDA Received Date07/25/2022
Supplement Report FDA Received Date09/05/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/18/2015
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BONE CEMENT (MFG. AND PRODUCT NAME UNKNOWN); SIGMA PS CEM FEM SZ2.5 R; SIGMA STAB GVF INS 2 12.5MM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient SexFemale
Patient Weight74 KG
-
-