Model Number ATG120204
Medical Device Problem Codes Deflation Problem (1149); Inflation Problem (1310); Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/27/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that prior to an angioplasty procedure, the balloon allegedly had abnormal inflation.It was further reported that the balloon allegedly had deflation problem.The procedure was completed using another device.There was no patient contact.
 
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history record is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.(expiry date: 02/2024).
 
Additional Manufacturer Narrative
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one atlas gold pta catheter was returned for evaluation.No specific anomalies were noted during the visual evaluation.In the functional testing, the device was inflated with an in-house presto inflation device, and the balloon inflated up to 16 atm without any issue and maintained pressure and shape.Further, the balloon took 20 seconds to deflate completely , which is within the acceptable range as per the product performance speciation.Therefore, the investigation into the reported inflation issue and material deformation was unconfirmed as the balloon inflated and maintained shape and pressure without any issues.The investigation into the reported deflation issue was unconfirmed as the balloon inflated and deflated without any issue and the deflation time was within the acceptable range as per the product performance specification.A definitive root cause for the reported inflation issue, material deformation and deflation issue could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 02/2024), g3.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that prior to an angioplasty procedure, the balloon allegedly had abnormal inflation.It was further reported that the balloon allegedly had deflation problem.The procedure was completed using another device.There was no patient contact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATLAS GOLD
Common Device Name
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
Manufacturer (Section G)
FUTUREMATRIX INTERVENTIONAL
1605 enterprise street
athens 75751
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15101363
Report Number2020394-2022-00605
Device Sequence Number5255465
Product Code DQY
UDI-Device Identifier00801741060946
UDI-Public(01)00801741060946
Combination Product (Y/N)N
Initial Reporter CountryIS
PMA/510(K) Number
K181323
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility,Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberATG120204
Device Catalogue NumberATG120204
Device Lot Number93PG0149
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/12/2022
Supplement Date Received by Manufacturer09/09/2022
Initial Report FDA Received Date07/26/2022
Supplement Report FDA Received Date09/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-