Consumer reported complaint for high blood glucose test results.The customer is concerned with test results from results obtained of 222 and 287 mg/dl; tests were performed back to back and customer did not use the proper back to back testing technique.The customer¿s expected pm fasting blood glucose test result range is 108-125 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call, a back to back blood test was not performed by the customer.The product is not stored according to specification (bathroom).The test strip lot manufacturer¿s expiration date is 12/22/2022 and open vial date is unknown.The customer did not have another vial of test strips that had been stored and handled correctly.The meter memory was reviewed for previous test result history: result 1: 222 mg/dl, date: (b)(6) 2022, time: 2:11 pm fasting.Result 2: 287 mg/dl, date: (b)(6) 2022, time: 2:14 pm fasting.Result 3: 125 mg/dl, date: (b)(6) 2022, time: 2:57 pm fasting.Result 4: 165 mg/dl, date: (b)(6) 2022, time: 2:38 pm fasting.Result 5: 116 mg/dl, date: (b)(6) 2022, time: 2:59 pm fasting.
|