| Model Number |
GLM915020 |
| Medical Device Problem Codes |
Stretched (1601); Failure to Advance (2524)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/06/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
As reported by the field, during a coil embolization, a 1.5mm x 2cm galaxy g3 mini coil (glm915020, 30624671) was being used as the third coil.The coil couldn't be advanced nor retrieved by the physician.The physician pulled the coil and the unspecified microcatheter (mc) at the same time.After inspection, it turns out to be stretched.The physician used a same like product.The procedure was successfully finished.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Information regarding patient weight, height, medical history, race, and ethnicity was not reported.Procode: krd/hcg.Initial reporter: the customer contact information, including name, occupation, phone, fax, and e-mail address, was not reported the device has been discarded; therefore, no further investigation can be performed.A manufacturing record evaluation (mre) was performed for the finished device 30624671 number, and no non-conformances related to the malfunction were identified.Missing information from this report is identified as blank; this information was not provided in the reported event or available at the time of report submission.The company is seeking this information through the event investigation.This report is being submitted pursuant to the provisions of 21 cfr, part 4.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by cerenovus, or its employees that the report constitutes an admission that the product, cerenovus, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint (b)(4).A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.Complaint conclusion: as reported by the field, during a coil embolization, a 1.5mm x 2cm galaxy g3 mini coil (glm915020, 30624671) was being used as the third coil.The coil couldn't be advanced nor retrieved by the physician.The physician pulled the coil and the unspecified microcatheter (mc) at the same time.After inspection it turns out to be stretched.The physician used a same like product.The procedure was successfully finished.No additional information is available.The device has been discarded; therefore, no further investigation can be performed.A manufacturing record evaluation (mre) was performed for the finished device 30624671 number, and no non-conformances related to the malfunction were identified.Impeded in microcatheter with no cerebral target loss and coil unraveled/stretched in microcatheter are known potential product failures associated with the use of the device.The instructions for use (ifu) warns that if the microcoil system becomes immobile in the infusion microcatheter, apply a gentle push-pull motion to free it.If unsuccessful, remove both microcatheter and microcoil system together as a unit and replace with new devices.Coil stretching is indicative of the application of excessive force, possibly in an attempt to overcome the reported resistance.With the information available and without the product available for analysis, the reported customer complaints could not be confirmed.Based on the manufacturing record evaluation, there is no indication that the events are related to the device manufacturing process.The exact cause of the event could not be determined; however, there are circumstances of the procedure that may have contributed to the reported failures.Multiple attempts to obtain additional information were unsuccessful.If information is provided at a later date, the file will be reopened and processed accordingly.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the events were related to a manufacturing or design issue, no corrective actions will be taken at this time.
|
| |
|
Search Alerts/Recalls
|
|