| Model Number |
DFT315 |
| Medical Device Problem Codes |
Difficult to Advance (2920); Scratched Material (3020)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
06/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A physician reported that after implantation of an intraocular lens (iol) the lens was unusually stuck in a folded condition for a longer period after it started to unfold there was a stretch mark in the middle.The surgery was completed but reserving the chance of explantation and re-implant with new lens.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
A sample device was not returned for analysis.The product history record could not be reviewed because facility did not provide a lot number or any identification traceable to the manufacturing documentation.Root cause has not been identified the manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided in h.3, h.6 and h.10.The provided photo shows a video screen displaying an image of the lens in the eye.What appears to be a linear mark is observed near the center of the optic.We are unable to make a determination from the photo.The account indicated the use of a qualified cartridge and viscoelastic.The associated handpiece was not provided.It is unknown if a qualified handpiece was used.Based on our observation of the attached photo, the lens appears to have a linear mark/scratch.The product was not returned to evaluate.It is difficult to make a final determination without evaluation of the physical sample.The manufacturer internal reference number is: (b)(4).
|
| |
|