| Model Number |
2130-02 |
| Medical Device Problem Code |
Overheating of Device (1437)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation on june 30, 2022 that a flexiva pulse id laser fiber was used during a procedure, performed on (b)(6) 2022.During the procedure, the laser fiber produced smoke when it was attached into the laser console.There were no patient complications reported as a result of this event.No further information has been obtained despite good faith efforts.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation on june 30, 2022 that a flexiva pulse id laser fiber was used during a procedure, performed on (b)(6) 2022.During the procedure, the laser fiber produced smoke when it was attached into the laser console.There were no patient complications reported as a result of this event.**additional information received on july 26, 2022** it was reported to boston scientific corporation on june 30, 2022, that a flexiva pulse id laser fiber was used during a holmium laser ablation procedure to treat stones in the bladder, ureter, and kidney, performed on (b)(6) 2022.The laser unit used was a lumenis pulse 100 watt laser console with laser settings of 0.3 joules and 53 hertz.During the procedure, the fiber connector was warm to the touch and smoke was observed.There were no patient complications reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h2: block b5 and block e: initial reporter updated based on the additional information received on july 26, 2022.Block h6: medical device problem code a1005 captures the reportable event of fiber connector overheats.Block h10: the device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation on june 30, 2022 that a flexiva pulse id laser fiber was used during a procedure, performed on (b)(6) 2022.During the procedure, the laser fiber produced smoke when it was attached into the laser console.There were no patient complications reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt, this product was thoroughly analyzed in our quality assurance laboratory.Visual analysis of the returned fiber confirmed the reported smoke produced by the fiber event as analysis identified a fracture within the connector.During functional testing, there was no light present from the sub miniature a (sma) connector, indicating a fracture within the connector.The fiber connector may have a developed a microfracture that with use or handling can become larger resulting in an insufficient aiming beam and low laser energy output and ultimately can lead the connector to become overheated.According to the device instructions for use.Do not bend fiber at sharp angles.Visible light seen leaking from the fiber indicates fiber damage.
|
| |
|