HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP; VENTRICULAR (ASSISST) BYPASS
|
Back to Search Results |
|
Model Number 1103 |
Device Problem
Pumping Problem (3016)
|
Patient Problem
Low Blood Pressure/ Hypotension (1914)
|
Event Date 02/26/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
Investigation of this event is pending and a supplemental report will be sent upon its completion.This information was received from the mechanical circulatory support product surveillance registry study.Concomitant medical products: 1688t, non-defib, lead, implanted: (b)(6) 2008.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the patient was admitted for a planned procedure video-assisted thoracoscopic surgery (vats) but the procedure was postponed due to heparin not being held.Five days later, the patient developed hypotension with ventricular assist device (vad) low flows.The vats procedure was postponed for a second time and a pulmonary artery catheter was placed.The vats procedure was successfully performed in the following days and the patient was discharged.The vad remains in use.No further patient complications have been reported as a result of this event.
|
|
Event Description
|
It was further reported that the patient had the vats procedure with upper lobe wedge resection/bullectomy and chemical and mechanical pleurodesis.The patient received a right erector spinae plane catheter for post-op analgesia.
|
|
Manufacturer Narrative
|
A supplemental report is being submitted for additional information.Additional information was received indicating that the patient had the vats procedure with upper lobe wedge resection/bullectomy and chemical and mechanical pleurodesis.The patient received a right erector spinae plane catheter for post-op analgesia; b5 updated.Investigation of this event is pending and a supplemental report will be sent upon its completion.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
A supplemental report is being submitted for investigation completion product event summary: the ventricular assist device (vad) was not returned for evaluation.The reported low power event was not confirmed via review of the controller log files since log files covering the reported event date were not available for analysis.Information provided by the site indicated that in addition to the low flow event the patient was admitted for a planned procedure video-assisted thoracoscopic surgery (vats) but the procedure was postponed due to heparin not being held.Five days later, the patient developed hypotension.The vats procedure was postponed for a second time and a pulmonary artery catheter was placed.The vats procedure was successfully performed in the following days and the patient was discharged.Based on the available information, the device may have caused or contributed to the reported event.Based on risk documentation, multiple factors may have contributed to the low flow event including but not limited to thrombus at the inflow cannula/outflow graft, constriction at the outflow graft, poor vad f illing, inappropriate pump rotational speed.Based on review of past adverse events for this patient, it was noted that the patient had a history of hypotension events.Possible clinical factors that may have contributed to this event include the patient¿s pre-existing history and related comorbidities, the progression of their underlying disease, issues related to the therapeutic use of anticoagulant and antiplatelet medications and the patient's complex post-operative course.There are possible patient, pharmacological and procedural factors that may have contributed to this event.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|