BAXTER HEALTHCARE CORPORATION FLEXICAP; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE Back to Search Results
Catalog Number 5C4456
Medical Device Problem Code Product Quality Problem (1506)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a minicap had no iodine; further described as ¿the dye had not released and noticed the dye had dried out¿.This was observed during use of the device of peritoneal dialysis therapy.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Additional Manufacturer Narrative
The event occurred on an unspecified date of june 2022.Should additional relevant information become available, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
Correction made to h6: investigation conclusions: replace d15 with d1102.Additional information added to h10: the patient used expired product.This was noticed after use due to dry iodine.The flexicap product should not be used if the iodine is dry per requirements of the patient information leaflet.The expiry date of the product should also be checked prior to use.Should additional relevant information become available, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
Correction made to b5: it was reported that a flexicap (previously submitted as minicap) had no iodine; further described as ¿the dye had not released and noticed the dye had dried out¿.Additional information was added to h4, h6 and h10.H4: the lot was manufactured in may 2021.H10: a batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The device was not received for evaluation; therefore, a device analysis could not be completed.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXICAP
Common Device Name
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - SWINFORD
foxford road
swinford
EI  
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key15103537
Report Number1416980-2022-03839
Device Sequence Number4588070
Product Code KDJ
UDI-Device Identifier00085412349145
UDI-Public(01)00085412349145
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K152129
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/28/2022
Device Catalogue Number5C4456
Device Lot Number21E04H25
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/30/2022
Supplement Date Received by Manufacturer07/29/2022
09/09/2022
Initial Report FDA Received Date07/26/2022
Supplement Report FDA Received Date08/15/2022
09/27/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age55 YR
Patient SexMale
-
-