| Model Number |
3000TFX |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Gradient Increase (1270)
|
| Health Effect - Clinical Code |
Heart Failure/Congestive Heart Failure (4446)
|
| Date of Event |
07/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional narratives: stenosis, which develops progressively over time, can be due to a number of issues.Additionally, there can be a number of potential known and unknown patient related contributing factors.Structural valve deterioration (svd), a common reason for bioprosthesis explant or reoperation, encompasses multiple failure modes, including calcific and non-calcific degeneration, dehiscence, cusp thickening or fibrosis, or a combination of these.Such failure modes, occurring singularly or concomitantly, may contribute to stenosis and/or regurgitation.Alternatively, nonstructural dysfunction (nsvd) may also play a role in the development of valvular stenosis.Stenosis is most commonly related to patient factors, and is not usually an indication of a device malfunction.The device was not returned for evaluation, as it remains implanted.The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient with a 19mm aortic valve is being evaluated for a valve-in-valve procedure after an implant duration of 4 years, 7 months due to severe stenosis and mild regurgitation.The patient presented with severe sob and acute diastolic chf.The patient has a history of as, chf, pulmonary htn, htn, hld, dm, breast carcinoma, and former smoker.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: additional narratives: updated b7 per new information received.
|
| |
|
Additional Manufacturer Narrative
|
|
H10:additional narratives updated d4, h4, and h6 per new information received the most likely cause is patient factors, including hld and dm.The device history record (dhr) was reviewed and showed that this device met all manufacturing and sterilization specifications for product release prior to distribution.No issues were identified that would have impacted this event.Edwards will continue to review and monitor all events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
|
| |
|