• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SYN POR HO FEM COM SZ 13; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SYN POR HO FEM COM SZ 13; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 71306113
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994)
Date of Event 05/29/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Bilateral patient.It was reported that, following an initial left r3-tha on (b)(6) 2010, plaintiff experienced left hip pain and popping episodes.Plaintiff underwent revision surgery on (b)(6) 2018, where the acetabulum was noted to be loose with erosion and broken screw heads.Metallosis, debris and fluid were observed.The femoral stem was retained.(cover under (b)(4)) on (b)(6) 2022, plaintiff experienced left hip pain and x-rays showed hardware failure with fracture through the taper-neck of femoral stem retained during the 1st revision surgery.Plaintiff underwent a 2nd revision surgery on (b)(6) 2022, where the stem was explanted.An osteotomy repair with kinamed cable and trochanteric 2-hole grip plate and screws were required.Plaintiff outcome is unknown.
 
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Additional Manufacturer Narrative
The device was not returned for evaluation but the reported event could be confirmed since the x-ray showed hardware failure with fracture through the neck of the left hip arthroplasty and proximal migration of the left femoral shaft.The clinical/medical investigation concluded that, with the information provided, the clinical root cause of the reported femoral stem fracture cannot be confirmed.The patient impact beyond the pain, revision, and expected transient post-op convalescence period cannot be determined.No further clinical assessment is warranted at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history for the part number over the past 12 months and for the batch number based on historical data of the device did not reveal similar events for the listed device.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include but not limited to traumatic injury or surgical technique.The contribution of the device to the reported event could be corroborated as the device was broken and it was required a revision surgery to treat this event.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYN POR HO FEM COM SZ 13
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15104364
Report Number1020279-2022-03468
Device Sequence Number17436929
Product Code LPH
UDI-Device Identifier03596010291974
UDI-Public03596010291974
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/22/2020
Device Model Number71306113
Device Catalogue Number71306113
Device Lot Number10EM06669A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/30/2022
Supplement Date Received by Manufacturer08/17/2022
Initial Report FDA Received Date07/26/2022
Supplement Report FDA Received Date08/18/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/25/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-