• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS IOL; INTRAOCULAR LENS Back to Search Results
Model Number PCB00
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/03/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Per regulation eu (b)(4) (general data protection regulation), patient identifiers were not collected or recorded and therefore are not available.If implanted, give date: not applicable, lens was not implanted.If explanted, give date: not applicable, lens was not implanted, therefore not explanted.Telephone number: (b)(6).Device evaluation: the device was not returned for evaluation.Therefore, a failure analysis of the complaint device could not be performed.Manufacturing record evaluation: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search revealed that one additional complaint for this order number has been received for a delivery and plunger rod issue which is unrelated to the reported issue.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the haptic came off prior to implantation and with no contact with the patient's eye.Through follow-up we learned that the issue occurred after removal from the box, during the implantation preparation.No further details were provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key15105541
Report Number3012236936-2022-01864
Device Sequence Number19862628
Product Code HQL
UDI-Device Identifier05050474558267
UDI-Public(01)05050474558267(17)240823
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 07/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/23/2024
Device Model NumberPCB00
Device Catalogue NumberPCB0000200
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/04/2022
Initial Report FDA Received Date07/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/23/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-