| Model Number |
N/A |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
Bone Fracture(s) (1870); Failure of Implant (1924); Pain (1994)
|
| Date of Event |
07/06/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a surgery was performed for a proximal tibia tumor with orthopedic salvage system implants.Postoperatively, the surgeon found that the anchor plug broke.There was bone formation seen on the side view and partially seen on the front view of x-rays.The patient was revised.Attempts have been made and no further information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Report source: foreign - japan.The device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.No devices or photographs were received; therefore, the condition of the components is unknown.The device history record was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4; b5; g3; g6; h1; h2; h3; h6; h11.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.The device history record was reviewed and no discrepancies relevant to the reported event were found.Journal article provided confirms that the anchor plug is fractured at the base of the anchor plug and during the revision they found out that tibia was fractured.It was reported that fell however the reason for fall is unknown.A definitive root cause cannot be determined.This complaint is confirmed via journal article notes provided.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated:.An investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that surgery was performed for proximal tibia tumor with oss.Approximately 9 months post operatively, the patient experienced discomfort and fell.The surgeon found that the anchor plug broke.Bone formation was seen on side view, but the bone formation was partially seen on front view from x-rays.Subsequently, a revision was performed on (b)(6) 2022.Tibial fracture was noted as well.
|
| |
|
Search Alerts/Recalls
|