| Model Number |
H7493942812450 |
| Medical Device Problem Codes |
Break (1069); Difficult to Remove (1528); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/03/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that shaft break occurred.The 90% stenosed target vessel was located in the moderately calcified and mildly tortuous left main artery.A 4.5mm x 12mm synergy megatron drug-eluting stent was successfully implanted, however, the stent delivery system was difficult to remove from the stent.Forceful removal was required and the catheter fractured.An attempt to remove the fractured catheter via snare was unsuccessful and the device had to be removed by a vascular surgeon using a cutdown of the brachial artery.The procedure was completed successfully.Hospital admission beyond standard of care was required, however, the patient is fully recovered.There were no further complications reported.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.The 90% stenosed target vessel was located in the moderately calcified and mildly tortuous left main artery.A 4.5mm x 12mm synergy megatron drug-eluting stent was successfully implanted, however, the stent delivery system was difficult to remove from the stent.Forceful removal was required and the catheter fractured.An attempt to remove the fractured catheter via snare was unsuccessful and the device had to be removed by a vascular surgeon using a cutdown of the brachial artery.The procedure was completed successfully.Hospital admission beyond standard of care was required, however, the patient is fully recovered.There were no further complications reported.It was further reported that the post deployment, balloon was inflated three times.A few seconds were waited for balloon deflation before pulling back the device.It was confirmed under fluoroscopy that the all the contrast had left the balloon prior to the withdrawal attempt.A 6f guide catheter was used during the procedure.No significant angulation or unusual take off was present.It was noted that the balloon would not come out of the stent.There was no difficulty advancing or positioning the stent at the lesion site.No other device was present in the vessel.The device could not maintain pressure, and no gradual loss of pressure was noted in the inflation device.No leaks were noted from the shaft of the device.No further patient complications were reported in relation to this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: a 4.50 x 12 mm synergy megatron mr tent delivery system (sds), was returned without the distal section of the device containing the inner/outer shaft polymer extrusion, balloon and tip.The device was returned without the distal shaft containing the stent.Distal shaft of the device not returned so the balloon cones could not be reviewed.Distal shaft of the device not returned so the tip of the device could not be reviewed.A visual and tactile examination of the hypotube shaft found multiple kinks along the length of the hypotube shaft.A visual and tactile examination of the outer and mid-shaft section found a break in the midshaft region at 119 cm distal to the distal end of the strain relief with the corewire exposed.No other issues were identified during analysis.
|
| |
|