• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. TROCAR INTRODUCER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. TROCAR INTRODUCER Back to Search Results
Model Number MAJ-1379
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 06/20/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The suspect device has not been returned to olympus for evaluation.The investigation is in process.Once the investigation has been completed, a supplemental report will be submitted with device evaluation results.
 
Event or Problem Description
The customer reported to olympus, near the end of the therapeutic laparoscopy, part of the subject device was damaged.The subject device was attached to olympus endoscope ltf-s190-5 and used as an insertion aid.A portable x-ray did not confirm any debris in abdominal cavity, no shadows seen.After the procedure was completed, a ct scan confirmed pieces fell into the abdominal cavity.A reoperation was performed (b)(6) 2022 and the foreign material was removed.It is unclear when c0r47 was used by the customer.The patient was discharged without further issues (b)(6) 2022.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide a correction, additional information from the customer, and the results of the device evaluation, legal manufacturer¿s investigation and the device history record (dhr) review.New information was added to the following fields: b5, d9, h3, h6, h10.Correction: e1 - hayata is the contact's last name only.The device was returned to olympus for evaluation and the customer's allegation was confirmed.Inspection of the device found the broken part was not part of the hook but was part of the holding parts at the opposite side.As a result of inspection of the fractured surface, a wavy pattern which was typical to fatigue failure was noted.The device history records (dhr) for this device could not be reviewed since the lot number was not provided.Olympus ships devices manufactured according to all applicable procedures and meet final product release criteria.A definitive root cause was not identified.Based on the available information, the legal manufacturer determined the probable cause of the failure is likely due to the combined effect of stress by the repeated use and stress.However, the cause of breakage was not specified since no more analysis with the actual device was available.Olympus will continue to monitor the field performance of the device.
 
Event or Problem Description
Additional information was received from the customer: the foreign material was removed from the abdominal cavity on (b)(4) 2022, using a laparoscope and grasping forceps under general anesthesia.The device information for the laparoscope and grasping forceps is unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TROCAR INTRODUCER
Common Device Name
TROCAR INTRODUCER
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer Contact
masaharu hirose
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8-520
JA   965-8520
426422891
MDR Report Key15106218
Report Number9610595-2022-00302
Device Sequence Number12055693
Product Code GCJ
UDI-Device Identifier04953170316159
UDI-Public04953170316159
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K201832
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,User Facility
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 09/20/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMAJ-1379
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/26/2022
Supplement Date Received by Manufacturer09/16/2022
Initial Report FDA Received Date07/26/2022
Supplement Report FDA Received Date09/20/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
LTF-S190-5(2824359).
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-