• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD MEDICAL (BD WEST) MEDICAL SURGICAL BD PRECISIONGLIDE¿ NEEDLE; HYPODERMIC SINGLE LUMEN NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL (BD WEST) MEDICAL SURGICAL BD PRECISIONGLIDE¿ NEEDLE; HYPODERMIC SINGLE LUMEN NEEDLE Back to Search Results
Model Number 305110
Medical Device Problem Code Partial Blockage (1065)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/06/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Investigation summary bd implemented a thorough investigation following our (capa 2166514) corrective preventative action protocol.This was opened for a trend in reports of ¿needle clogged / blocked¿, ¿aspirate / draw (cannot / difficult)¿, and ¿needle bent¿.All three of these issues have the same net effect: temporary interruption of flow of insulin into the reservoir.Based on the investigation, these issues appear to be related to the use case which involves drawing insulin from a vial into a syringe and then filling the reservoir of the insulin pump via a small port off the body in an iterative process, utilizing existing / ¿off the shelf¿ products rather than a product designed for this use.Bd products are designed to be single use, and because this is an "off the shelf" product, bd specifications do not take into account any requirements specific to the tandem use case.The investigation concludes the complaints are related to the use case and not due to a product nonconformance.
 
Event or Problem Description
It was reported that bd precisionglide¿ needles had fill resistance issue.The following information was provided by the initial reporter: it was reported by the customer that they encountered a "fill resistance issue.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD PRECISIONGLIDE¿ NEEDLE
Common Device Name
HYPODERMIC SINGLE LUMEN NEEDLE
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15106959
Report Number1911916-2022-00369
Device Sequence Number4213490
Product Code FMI
UDI-Device Identifier30382903051107
UDI-Public30382903051107
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K021475
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date04/30/2025
Device Model Number305110
Device Catalogue Number305110
Device Lot Number0072110
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/27/2022
Initial Report FDA Received Date07/26/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/12/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age58 YR
Patient SexFemale
-
-