| Model Number |
DL-344-10-K |
| Medical Device Problem Codes |
Signal Artifact/Noise (1036); Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/03/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4) the explanted device was returned to neuropace and investigated.Investigation of the explanted lead suggests that the lead may have been damaged as a result of a fixation method.Review of the lead data also confirmed the appearance of signals and high impedance indicative of a probable lead fracture.
|
| |
|
Event or Problem Description
|
|
During patient ecog review on (b)(6) 2022, the doctor observed an impedance alert on one of the depth leads, indicating insufficient charge and a possible lead break.Troubleshooting included; repeated impedance checks, palpation of the lead entry.Stimulation was disabled to that lead.There is no information suggesting the cause of the break.The lead was explanted on (b)(6) 2022 and returned to neuropace for investigation.
|
| |
|