HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP; VENTRICULAR (ASSISST) BYPASS
|
Back to Search Results |
|
Model Number 1103 |
Device Problem
Pumping Problem (3016)
|
Patient Problem
Bacterial Infection (1735)
|
Event Date 07/21/2022 |
Event Type
Injury
|
Event Description
|
It was reported that the patient was hospitalized with bacteremia and subtherapeutic international normalized ratio (inr).The patient was given fluids for lower flows and heparin for the subtherapeutic inr.Treatment for the bacteremia was unknown.It was further noted that flows were stable in the 2.9-3.0 range with good pulsatility.A computerized tomography (ct) scan showed fluid collection beneath the xiphoid process and sternum.The ventricular assist device (vad) remains in use.No further patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Investigation of this event is pending and a supplemental report will be sent upon its completion.Additional information has been requested regarding further event details, but it was not available at the time of this report.If additional information is received, the event will be updated and a supplemental report will be sent.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was further reported that the patient went to the operating room for a subxiphoid washout with chest tube placement and a wound vacuum assisted closure (vac) placement.An "open/free" pocket was found under the xiphoid that was filled with fluid.Fluid collection was believed to be the source of infection.
|
|
Manufacturer Narrative
|
A supplemental report is being submitted for additional event details.B5 describe event or problem was updated.Investigation of this event is pending and a supplemental report will be sent upon its completion.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
A supplemental report is being submitted for device evaluation.Product event summary: (b)(6) was not returned for evaluation.The reported low flow event was confirmed via review of the controller log files which revealed a slight decrease in power consumption and estimated flows starting on 18/jul/2022.Two (2) low flow alarms were logged on 20/jul/2022.Information provided by the site indicated that the patient was hospitalized with bacteremia and subtherapeutic international normalized ratio (inr).The patient was given fluids for lower flows and heparin for the subtherapeutic inr.Treatment for the bacteremia was unknown.A computerized tomography (ct) scan showed fluid collection beneath the xiphoid process and sternum.The patient went to the operating room for a subxiphoid washout with chest tube placement and a wound vacuum assisted closure (vac) placement.An "open/free" pocket was found under the xiphoid that was filled with fluid.Fluid collection was believed to be the source of infection.Based on the limited information available, the device may have caused or contributed to the reported event.Based on the risk documentation, possible causes of the reported low flow event may be attributed to multiple factors including but not limited to thrombus at the inflow cannula/outflow graft, constriction at the outflow graft, inappropriate pump rotational speed, and/or poor vad filling.Per the instructions for use, infection is a known potential complication associated with the implantation of a vad.Based on review of past adverse events for this patient, it was noted that the patient did not have a history of infection events.Possible clinical factors that may have contributed to this event include the patient¿s pre-existing history and related comorbidities, the progression of their underlying disease, and the patient's complex post-operative course.There are possible patient, pharmacological and procedural factors that may have contributed to this event.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|