MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ARROW OUS PI PICC KIT: 1-L 4.5FR X 50CM CG+; CATHETER,INTRAVASCULAR,THERAP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ARROW OUS PI PICC KIT: 1-L 4.5FR X 50CM CG+; CATHETER,INTRAVASCULAR,THERAP Back to Search Results
Model Number IPN914728
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 04/21/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that "after implantation, at the follow-up x-ray, there was a presence in the lung that resulted from the x-ray.It was decided to retrieve the foreign body, which turned out to be the guide wire sheath that was located in the pulmonary artery.The patient is fine and the picc was left in situ.".It was reported the foreign body was retrieved by vascular surgery and "a catheter with a loop at the tip was introduced, and under radiographic control the foreign body was completely removed".
 
Additional Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as it was reported that the sample is not available for return.A device history record review was performed and no relevant findings were identified.Without the device to evaluate the complaint could not be confirmed and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Event or Problem Description
It was reported that "after implantation, at the follow-up x-ray, there was a presence in the lung that resulted from the x-ray.It was decided to retrieve the foreign body, which turned out to be the guide wire sheath that was located in the pulmonary artery.The patient is fine and the picc was left in situ.".It was reported the foreign body was retrieved by vascular surgery and "a catheter with a loop at the tip was introduced, and under radiographic control the foreign body was completely removed".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW OUS PI PICC KIT: 1-L 4.5FR X 50CM CG+
Common Device Name
CATHETER,INTRAVASCULAR,THERAP
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL DE MEXICO S.A. DE C.V.
ave. washington 3701
colonia panamericana, chihuahua
chihuahua 31200
MX   31200
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key15107665
Report Number9680794-2022-00457
Device Sequence Number4214093
Product Code LJS
UDI-Device Identifier10801902157543
UDI-Public10801902157543
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K112896
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 07/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Expiration Date02/28/2023
Device Model NumberIPN914728
Device Catalogue NumberCDA-45041-HPK1A
Device Lot Number13F21H0557
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/06/2022
Supplement Date Received by Manufacturer08/22/2022
Initial Report FDA Received Date07/26/2022
Supplement Report FDA Received Date08/24/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/24/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;