SMITH & NEPHEW, INC. JOURN ART INS BCS STD 5-6 RT 9; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Model Number |
74023251 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
07/07/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, after total knee arthroplasty was performed on an unknown date, the patient experienced instability and felt that something was floating in their knee.A computed tomography showed a possible piece in the knee joint.This adverse event was treated with revision surgery on (b)(6) 2022, in which the poly was explanted and replaced with a journey i revision poly; in addition, the broken piece was retrieved.Patient's current health status is unknown.
|
| |
|
Event or Problem Description
|
|
It was reported that, after total knee arthroplasty was performed on (b)(6) 2009, the patient experienced instability and felt that something was floating in their knee.A computed tomography showed a possible piece in the knee joint.This adverse event was treated with revision surgery on (b)(6) 2022, in which the poly was explanted and replaced with a journey i revision poly; in addition, the broken piece was retrieved.Patient's current health status is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the device was not returned for evaluation but the pictures were reviewed, and revealed that the insert is broken.The clinical/medical investigation concluded that the provided primary implant record and photos of the explanted insert were assessed; however, no clinical factors were noted which would have contributed to the event.The patient's current health status, medical history and comorbidities are unknown.The patient impact beyond that which was reported could not be determined.No further medical assessment is warranted at this time.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for knee systems revealed that the implant can break or become damaged as a result of strenuous activity or trauma, and has a finite expected service life and may need to be replaced in the future.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes for this event could include but not limited to traumatic injury or surgical technique.The contribution of the device to the reported event could be corroborated as the device was damaged and it was required a revision surgery to treat this event.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|