| Catalog Number |
PXSLIMLAN150T45 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/05/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has been returned and the investigation results are pending.A follow up mdr will be submitted upon completion of the device investigation.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a coil embolization procedure in the subclavian collateral artery using a lantern delivery microcatheter (lantern), a non-penumbra guide catheter.During the procedure, after initially advancing the lantern through the guide catheter, the physician fractured the lantern mid-shaft.Therefore, the lantern was removed.The procedure was completed using a new lantern.There was no report of an adverse effect to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation of the returned lantern confirmed a fracture on its mid-shaft and revealed a kink near the fractured location.If the lantern is advanced at an extreme angle during use, damage such as a kink and fracture may occur.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
| |
|