• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON MEDICAL (SINGAPORE) BD CATHENA SAFETY IV CATHETER; INTRAVASCULAR CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON MEDICAL (SINGAPORE) BD CATHENA SAFETY IV CATHETER; INTRAVASCULAR CATHETER Back to Search Results
Model Number 386803
Medical Device Problem Codes Retraction Problem (1536); Defective Component (2292)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/27/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported while using bd cathena safety iv catheter the safety mechanism failed.There was no report of patient impact.The following information was provided by the initial reporter: placed on a patient, when the needle is withdrawn, the safety device is not engaged.
 
Additional Manufacturer Narrative
A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
 
Additional Manufacturer Narrative
H6: investigation summary no photos or samples were received by our quality team for evaluation therefore the failure mode could not be verified, and the root cause could not be determined.A device history record could not be evaluated as the lot number is unknown.
 
Event or Problem Description
It was reported while using bd cathena safety iv catheter the safety mechanism failed.There was no report of patient impact.The following information was provided by the initial reporter: placed on a patient, when the needle is withdrawn, the safety device is not engaged :.
 
Event or Problem Description
It was reported while using bd cathena safety iv catheter the safety mechanism failed.There was no report of patient impact.The following information was provided by the initial reporter: placed on a patient, when the needle is withdrawn, the safety device is not engaged :.
 
Additional Manufacturer Narrative
The following field has been corrected: h.6.Imdrf annex a code: a0510 h3 other text : see h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD CATHENA SAFETY IV CATHETER
Common Device Name
INTRAVASCULAR CATHETER
Manufacturer (Section D)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer (Section G)
BECTON DICKINSON MEDICAL (SINGAPORE)
30 tuas avenue 2
singapore
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8015296192
MDR Report Key15108735
Report Number8041187-2022-00395
Device Sequence Number14506625
Product Code FOZ
UDI-Device Identifier00382903868032
UDI-Public00382903868032
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K172506
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number386803
Device Catalogue Number386803
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/27/2022
Supplement Date Received by Manufacturer08/04/2022
09/13/2022
Initial Report FDA Received Date07/26/2022
Supplement Report FDA Received Date08/31/2022
09/15/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-