| Model Number |
VCP398H |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a patient underwent an unknown procedure in 2022 and suture was used.During the procedure, it was noted that the suture cannot be used because at the first step on the tissue the needle detached from the wire the stitch has been pulled out and a new one entered.This cause a slight delay.No adverse patient consequences were reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information was requested, and the following was obtained: was there any adverse consequence associated with the patient? no.Was there any change in the patient¿s post-operative care due to the prolonged procedure? no.Were there any unexpected outcomes or complications as a result of the prolonged surgery time? none.How long (in minutes) did the surgery prolong? few minutes, time needed to take a new suture.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This is a combination product, and the event has been reviewed for both the suture and the triclosan.Trade name : irgacare.Active ingredient(s): triclosan.Dosage form : suture/solid/parenteral.Strength : = 275 g /m.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 08/15/2022.This is a combination product, and the event has been reviewed for both the suture and the triclosan.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc., or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Component code: g07002 no problem detected.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.The product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that one opened box with packets of product were returned with the packaging closed.Upon initial inspection, of the samples, no external damages were observed on the packets.The visual inspection was performed on samples, the packets were opened, and no defects were detected.The swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along the strand to detect any issues related and no defects were observed during evaluation.A functional test was performed, and the pull force result was above the minimum requirements.The device performed without any defect noted.As part of our quality process, the manufacturing records of this lot-serial number were reviewed, and the manufacturing standards were met prior to the release of this batch.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Trade name irgacare.Active ingredient(s) triclosan.Dosage form suture/solid/parenteral.Strength 275 g /m.
|
| |
|
Search Alerts/Recalls
|
|
|