| Model Number |
7205459 |
| Medical Device Problem Code |
Flaked (1246)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during a knee arthroscopy, metallic microparticles were found inside the joint caused by the incisor plus shaver blade; the particles were aspirated along with the serum.The procedure was successfully completed with no surgical delay using the same device.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference: (b)(4).H3, h6: the reported device was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found no similar reported events.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Per case details, metallic microparticles were found inside the joint caused by the shaver blade.Reportedly, the particles and serum were removed by aspiration with ¿no damaged reported,¿ or patient injury alleged.¿the patient¿s health condition is stable.¿ no relevant supporting clinical information has been provided to assist with a clinical investigation.Therefore, based on insufficient information, a thorough clinical assessment cannot be performed at this time and the patient impact beyond the reported events could not be determined.Should any additional clinical information be provided, this complaint could be re-evaluated.There was no relationship found between the device and the reported event.The complaint was not confirmed.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.Correction in h6 (health effect - impact code).
|
| |
|
Additional Manufacturer Narrative
|
|
H10: internal complaint reference:(b)(4).H2: additional information in b5 & b7.Correction in h6 (health effect - impact code).
|
| |
|
Event or Problem Description
|
|
It was reported that during a knee arthroscopy, metallic microparticles were found inside the joint caused by the incisor plus shaver blade; the particles were aspirated along with the serum.The procedure was successfully completed with no surgical delay using the same device.No further complications were reported.Patient status is stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).The reported device was received for evaluation.There was a relationship found between the device and the reported event.A visual evaluation of the device showed that it was not returned in the original packaging.The color schematics match the print.The teeth on both the inner and outer blades appear sharp and aligned.Removal of the inner blade showed three bands around the circumference of the shaft.Each band had grooves worn into them.There is at least one band on the inside circumference of the outer blade, just below the blade window.No particles fell out when the two sections were separated.A functional evaluation of the device showed that there was no drag or resistance felt when turning the blade manually.There were no particles when the device was run on a control system.The device functioned as intended in all modes.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review found no similar reported events.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.No relevant supporting clinical information has been provided to assist with a clinical investigation.Therefore, based on insufficient information, a thorough clinical assessment cannot be performed at this time and the patient impact beyond the reported events could not be determined.Should any additional clinical information be provided, this complaint could be re-evaluated.The complaint was confirmed and the root cause was associated with component failure.Factor that may have contributed to the reported event is insufficient irrigation during use.No containment or corrective actions are recommended at this time.
|
| |
|