ANGIOMED GMBH & CO. MEDIZINTECHNIK KG LIFESTENT 5F VASCULAR STENT Back to Search Results
Model Number 5F051503CS
Medical Device Problem Codes Material Deformation (2976); Positioning Problem (3009)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/29/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a stent placement procedure, the stent allegedly had a positioning problem.It was further reported that catheter allegedly buckled.There was no reported patient injury.
 
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history records will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiry date: 03/2024).
 
Additional Manufacturer Narrative
H10: manufacturing review: the lot history records of this lot were reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met the specification prior to shipment.Investigation summary: the sample was returned for evaluation.On the returned sample the outer stability sheath was found with compressive crinkles which may indicate insertion/ crossing difficulty.Images/ movie demonstrating crossing failure were not provided, however the provided information about the event and the compressive crinkles found lead to confirmed result for failure to advance leading to catheter deformation.In this case stent deployment was attempted via pedal access, the vessel was not tortuous but heavily calcified; the guidewire was running smoothly inside the system, and the lesion was not pre dilated; system compatible introducer and guidewire size were used for access.Based on the investigation of the provided information, the investigation is confirmed for failure to cross the lesion, and subsequent buckling.A definite root cause for the reported event could not be determined.Labeling review: in reviewing the relevant labeling for this product the potential issue was found addressed.The instruction for use state: 'if resistance is met during stent system introduction, the stent system should be withdrawn and another stent system should be used.', and 'pre-dilatation of the lesion with a balloon dilatation catheter is recommended.' in regards to required materials and access the instruction for use states: 'gain ipsilateral or contralateral femoral access utilizing an appropriate introducer sheath ¿ 5f (1.67 mm) or larger introducer sheath.(.) insert a guidewire of appropriate length and 0.014 inch (0.36 mm) - 0.035 inch (0.89 mm) diameter across the lesion to be stented via the introducer sheath.' h10: d4 (expiry date: 03/2024), g3 h11: h6(method, result, conclusion) h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that during a stent placement procedure, the stent allegedly had a positioning problem.It was further reported that catheter allegedly buckled.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFESTENT 5F VASCULAR STENT
Common Device Name
VASCULAR STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
Manufacturer (Section G)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM   76227
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key15110782
Report Number9681442-2022-00223
Device Sequence Number4116289
Product Code NIP
UDI-Device Identifier00801741120664
UDI-Public(01)00801741120664
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
P070014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5F051503CS
Device Catalogue Number5F051503CS
Device Lot NumberANFQ2158
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/29/2022
Supplement Date Received by Manufacturer09/01/2022
Initial Report FDA Received Date07/27/2022
Supplement Report FDA Received Date09/03/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-