| Model Number |
332097 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
05/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
(b)(6) reported by the user facility: customer reports that the obstetric patient had a contraction during placement of the epidural catheter.The anesthesiologist believes the tip of the catheter sheared off during the contraction.The patient has had an mri which shows metallic artifact.They are leaving the tip in place for now and will continue to monitor her.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).No sample was provided for evaluation.Based on the data from the investigation we are unable to determine the root cause of the reported incident.The reported defect was unable to be confirmed.The actual defective device is a valuable tool in investigating the cause of this incident.It should be noted that ifu states, "never pull the catheter back through the epidural needle due to the possibility of shearing or kinking the catheter.When removing the catheter, excessive force should never be applied.If the catheter becomes difficult to withdraw, consult standard textbooks for specific techniques." review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|
| |
|
Search Alerts/Recalls
|