U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION STONETOME; DISLODGER, STONE, BILIARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION STONETOME; DISLODGER, STONE, BILIARY Back to Search Results
Model Number M00535150
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/07/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).(b)(4).The device has not been received for analysis.Upon receipt and completion of the device analysis, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation that a stonetome was used in the papilla during a papillary incision procedure performed on (b)(6) 2022.During the procedure, the cutting wire of the stonetome broke.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
Block e (initial reporter city, state):(b)(6) block h6: medical device problem code a0401 captures the reportable event of cutting wire broken.Block h10: the returned stonetome was analyzed, and a visual evaluation noted that the cutting wire was broken at 4mm from the proximal pierced hole and also it was kinked.The device was observed under magnification and the cutting wire was broken, kinked from the proximal pierced hole, and blackened.No other problems with the device were noted.The reported event of cutting wire break was confirmed.Upon analysis, it was found that the cutting wire was broken, blackened, and kinked.The cutting wire being blackened indicates that the device was energized.Based on the condition of the device, the problem found could have been generated if the cutting wire was preactivated which may cause premature cutting wire fatigue and may compromise the cutting wire's integrity.The problems could have been generated also when energization was applied to the device constantly, or when it was activated in a incorrect position, being in contact with other medical devices.Once the cutting wire breaks, any attempt to remove the device from the scope can lead to the broken section hitting the working channel of the scope.This may lead to a kink in the cutting wire and tear the distal pierce hole.Based on all gathered information, the most probable root cause of this complaint is adverse event related to procedure.A review of the manufacturing documentation for this device revealed that no anomalies or deviations related to the event occurred during manufacturing.A product labeling review identified that the device was used per the directions for use (dfu) / product label.
 
Event or Problem Description
It was reported to boston scientific corporation that a stonetome was used in the papilla during a papillary incision procedure performed on (b)(6) 2022.During the procedure, the cutting wire of the stonetome broke.It was reported that no part of the cutting wire detached and fell into the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STONETOME
Common Device Name
DISLODGER, STONE, BILIARY
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key15112171
Report Number3005099803-2022-04051
Device Sequence Number4587358
Product Code LQR
UDI-Device Identifier08714729146667
UDI-Public08714729146667
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K946358
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/31/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/03/2024
Device Model NumberM00535150
Device Catalogue Number3515
Device Lot Number0028617167
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/07/2022
Supplement Date Received by Manufacturer08/11/2022
Initial Report FDA Received Date07/27/2022
Supplement Report FDA Received Date08/31/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/03/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-