MEDICAL COMPONENTS, INC. 2.6F X 50CM DUAL VASCU-PICC Back to Search Results
Model Number MD17012601
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device in this complaint was not returned for evaluation and no photographs were provided.The contract manufacturer conducted a review of the manufacture records for the lot number reported.Their investigation revealed the device was manufactured and inspected according to specification with no non-conformances or abnormalities.The manufacture process includes a 100% leak test performed as a last step in the process.This test would have detected any leaks, holes, or weak spots if it had existed at the time of manufacture.No issues or leaks were noted for this lot.The device was in use for more than one month prior to the event with no reported issues.Without an evaluation of the device a root cause cannot be determined.The instructions for use (ifu) contains the following warnings and precautions: do not use infusion equipment which can exceed the working pressure of 1.0bar max/750mmhg (14.5 psi).Only use infusion equipment complying with standards, which do not exceed a shut-off pressure of 1.0bar.Bolus injections should be slow and must not exceed the maximum bolus pressure of 1.2bar/900mmhg (17.4 psi).Small syringes will generate excessive pressure and may damage the catheter.The use of 10cc or larger syringes are recommended.Do not use sharp instruments near the extension lines or catheter lumen.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Information obtained from medwatch (b)(4) forwarded to medcomp.Line was documented as having blood and fluid under the dressing and was not flushing well.When the dressing was taken down the bedside nurse identified that the was leaking and the team decided to remove the line and just move to piv instead of replacing the picc.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.6F X 50CM DUAL VASCU-PICC
Common Device Name
VASCU-PICC
Manufacturer (Section D)
MEDICAL COMPONENTS, INC.
1499 delp drive
harleysville PA 19438
Manufacturer (Section G)
MEDICAL COMPONENTS, INC.
1499 delp drive
harleysville PA 19438
Manufacturer Contact
lynn winkler
1499 delp drive
harleysville, PA 19438
2152564201
MDR Report Key15112189
Report Number2518902-2022-00046
Device Sequence Number11909118
Product Code LJS
UDI-Device Identifier00884908146493
UDI-Public884908146493
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K102966
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/01/2023
Device Model NumberMD17012601
Device Catalogue NumberMD17012601
Device Lot NumberMQDZ810
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/13/2022
Initial Report FDA Received Date07/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/24/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age6 YR
Patient SexFemale
Patient Weight12 KG
-
-