Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was reported to be available for evaluation but has not yet been returned.If the device is received the investigation will be performed and a supplemental report will be submitted.The instructions for use (ifu) identifies premature or difficult coil detachment as potential complications associated with use of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that during delivery of the coil, the coil unintentionally detached inside the patient before they were able to reach the treatment site.The coil was removed from the patient using a snare device.The procedure was continued with additional coils with no further issues.The patient was reported to be in stable condition, and procedure outcome was good.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation of the returned coil system found the implant stretched, damaged, and separated from the pusher.The pusher was not returned for evaluation.The investigation found the implant's monofilament with a tensile break shape at the tip which is consistent with the device experiencing excessive force that exceeded the strength of the monofilament causing the implant to separate from the pusher.The physical evaluation of the device could not identify the conditions or circumstances that led to the damage, but the damage is consistent with the device experiencing forces over specification.
|
| |
|
Search Alerts/Recalls
|