Investigation: visual inspection: the following observations were made during the visual inspection: no abnormalities detected.Connected with rm 4260_1 (m.Blue).Permeability test: the test showed that the progav 2.0 is permeable.Computer control test: the computer controlled test showed the opening pressure of the progav 2.0, at a reference flow rate of 20 ml/h in a horizontal position of the valve, to be 8.33 cmh2o.This is within the specified tolerance of 8 cmh2o ± 3 cmh2o.Adjustability test: the progav 2.0 was found to be adjustable to all pressure settings.Braking force and brake function test: the braking force of the progav 2.0 was within the specified tolerance and the brake function operated as expected.Internal inspection of product: in order to verify whether the investigated valve was compromised by the known risks of hydrocephalus therapy, e.G.By a build-up of natural substances (protein, blood, or tissue particles) in the cerebrospinal fluid, the valve is finally opened with a special tool.After dismantling of the valve, deposits were found in progav 2.0.Results: we are obliged to inform you that the condition of the product could be affected due to the interval between the explantation date and the receipt of the product, because fungus and mold can develop.For the sake of thoroughness and transparency, we would like to point out that an adjustment test and permeability test were already carried out after the revision at the hospital.Based on our investigation results, we can determine deposits.The deposits had no effect to the technical properties at the time of the investigation.The cause of the aforementioned functional impairment is not known to us at the time of the examination.Proteins in the cerebrospinal fluid can influence the function temporarily and are known side effects in hydrocephalus therapy.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.From our point of view, no further regulatory actions are required.
|