• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US SROM NRHFEM W/PIN SM LFT 66X62; LPS AND S-ROM : KNEE FEMORAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US SROM NRHFEM W/PIN SM LFT 66X62; LPS AND S-ROM : KNEE FEMORAL Back to Search Results
Model Number 62-3411L
Medical Device Problem Code Loss of or Failure to Bond (1068)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 07/11/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery: (b)(6) 2009 patient (r) side revision of competitor products (unknown implants revised).Revision surgery: (b)(6)2010 patient (l) side revision of competitor products (unknown implants revised).Revision surgery: (b)(6) 2022 patient (l) revision of depuy products: femoral component loose and needed revision.Implants insitu (affected items listed following under 8.3.2010 - refer to source data).Femoral components removed and revised to larger stem and sleeve.Existing plateau re used.Tibia remains insitu.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
Additional information received indicated that there was no surgical delay reported.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: the device associated with this report was not returned.All available x-ray evidence were reviewed and cannot be establish the loosening condition reported.Based on the information, a defect or malfunction cannot be determined which could be attributed to the reported event, therefore the allegation is not confirmed.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).Where the product and lot code was provided, a manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SROM NRHFEM W/PIN SM LFT 66X62
Common Device Name
LPS AND S-ROM : KNEE FEMORAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
DEPUY RAYNHAM, A DIV. OF DEPUY ORTHO 1219655
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
700 orthpaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key15112741
Report Number1818910-2022-14210
Device Sequence Number14850296
Product Code KRO
UDI-Device Identifier10603295179481
UDI-Public10603295179481
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K896048
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2022
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/27/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number62-3411L
Device Catalogue Number623411L
Device Lot Number238411LA
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/11/2022
Supplement Date Received by Manufacturer08/10/2022
10/06/2022
Initial Report FDA Received Date07/27/2022
Supplement Report FDA Received Date08/16/2022
10/07/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/13/2008
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SROM KNEE FEM SLEEVE POR 34MM.; SROM NRH REPLC HINGE BRG QTY2.; SROM NRH TIB PLAT ASSY SM 21MM.; SROM NRH TIB SLEEVE POR 45MM.; SROM NRHFEM W/PIN SM LFT 66X62.; SROM* BASE TIB MOD UNTEXT MED.; SROM*DOME PAT UHMWPE 38MM.; SROM*EXTENSN,TIB STEM,11X100MM.; SROM*EXTNSN,FEM.STM,STR,15X100.
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-