Blank fields on this form indicate the information is unchanged, unknown, or unavailable.Investigation ¿ evaluation: shanghai huirong med.Science (china) contacted cook on 20jul2022 concerning the needle in a quick-core coaxial biopsy needle set (rpn: qcs-18-15.0-20t; lot# 14425881).It was reported that on 18jul2022 a 49-year-old female patient, weighing 56 kg underwent a lung puncture.During the procedure, the user had difficulty separating the inner cannula of the dtn needle provided in this set.The needle was replaced with a new device and the procedure was completed.Tools had to be used to separate the inner cannula of the dtn.No adverse effects were reported due to this occurrence.Reviews of the complaint history, device history record (dhr), instructions for use (ifu), quality control procedures, and specifications of the device, as well as a visual inspection and functional test of the returned device, were conducted during the investigation.Cook received one prior to use device and fourteen sealed devices.Resistance was noted with four of the sealed devices, but the inner cannula was able to be separated.With the prior to use device, tools were needed to separate the inner cannula of the dtn.The reported failure was confirmed on the prior to use device and the four sealed devices.Measurements of the inner and outer diameters of the devices were taken and found to be out of specification.Additionally, a document-based investigation evaluation was performed.In response to this incident, cook completed a review of the product device master record (dmr) and concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.Cook also reviewed the device history record (dhr).The dhrs for lot 14425881 and dtn sub-assembly lot, sa13770232, record no relevant non-conformances or other complaints.The information provided upon review of the returned device suggests that the device was manufactured out of specification.There is no evidence of non-conforming material in house or in the field.Cook reviewed product labeling.This product is supplied with an instructions for use (ifu) pamphlet, t_qc_rev6.In the instructions for use section, it states: 1.) if using the coaxial needle, insert the coaxial needle to the border of the lesion.In the how supplied section it states: upon removal from package, inspect the product to ensure no damage has occurred.Based on the information provided, inspection of the returned device, and the results of the investigation, cook determined the cause of this event is related to a quality control deficiency.A defect awareness was sent to the appropriate department and has been acknowledged and signed.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|