Model Number N/A
Medical Device Problem Code Difficult to Remove (1528)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
It was reported that difficulty was experienced when attempting to remove the trocar from a quick-core coaxial biopsy needle set prior to a supraclavicular lymph node puncture on (b)(6) 2022.The puncture needle was replaced to complete the procedure successfully.The patient did not experience any adverse effects due to this occurrence.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Investigation ¿ evaluation (b)(6) hospital (china) contacted cook on 15jul2022 concerning a quick-core coaxial biopsy needle set (rpn: qcs-18-15.0-20t; lot# 14434239).The customer reported that on (b)(6) 2022 an unknown patient was undergoing a supraclavicular lymph node puncture.The stylet of the trocar needle was difficult to remove during the procedure.The needle was replaced, and the procedure was completed.The patient did not experience any adverse effects due to this occurrence.Reviews of the complaint history, device history record (dhr), instructions for use (ifu), and quality control procedures of the device, as well as a visual inspection and functional test, were conducted during the investigation.One prior to use device was received by cook for evaluation.The inner cannula was attached rather tightly.Difficult removal was confirmed for this failure.Cook determined the product out of specification.Additionally, a document-based investigation evaluation was performed.In response to this incident, cook completed a review of the product device master record (dmr) and concluded that sufficient inspection activities are in place to identify this failure mode prior to distribution.The dhr was reviewed.The dhrs for lot# 14434239 and subassembly lots sa13770232 (dtn) and sa13770212 (dtn) record no relevant non-conformances.There is one other complaint on this lot for this same failure where there was difficulty removing the stylet from the dtn needle (coaxial needle).Based on this information, there is no evidence of non-conforming material in house or in the field.Cook also reviewed product labeling.This product is supplied with an instructions for use (ifu) pamphlet, t_qc_rev6.In the how supplied section it states: upon removal from package, inspect the product to ensure no damage has occurred.Based on the information provided, inspection of the returned device, and the results of the investigation, it was determined the cause of this event is related to a quality control deficiency.The failure was confirmed in lab testing.There are procedures and quality control steps for this non-conformance.A defect awareness was sent to the appropriate department which has been acknowledged and signed.The appropriate personnel have been notified.Cook will continue to monitor for similar complaints.Per the risk assessment no further action is required.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICK-CORE COAXIAL BIOPSY NEEDLE SET
Common Device Name
KNW INSTRUMENT, BIOPSY
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key15113450
Report Number1820334-2022-01285
Device Sequence Number14994446
Product Code KNW
UDI-Device Identifier00827002087903
UDI-Public(01)00827002087903(17)261221(10)14434239
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K973565
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue NumberQCS-18-15.0-20T
Device Lot Number14434239
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/15/2022
Supplement Date Received by Manufacturer11/29/2022
Initial Report FDA Received Date07/27/2022
Supplement Report FDA Received Date12/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/21/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age71 YR
Patient SexMale
Patient Weight65 KG
-
-