| Model Number |
775659 |
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/25/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer reported that there are air bubbles in the tube which is blocking the tube from functioning and making the machine ring.No injuries occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation is currently underway.Upon completion, the results will be forwarded.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record was reviewed and indicated that the product was released accomplishing all quality standards.The device was manufactured on january 27, 2022.Both photos and a physical sample were received for investigation.A visual and functional inspection was performed, and the reported condition could not be duplicated in the sample received.However, a review of the photos provided did confirm the reported condition.A corrective and preventative action has been initiated to address the reported issue.
|
| |
|
Search Alerts/Recalls
|