• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON LABORATORIES IRELAND LTD. ACRYSOF IQ ASPHERIC IOL WITH THE ULTRASERT PRE-LOADED DELIVERY SYSTEM; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON LABORATORIES IRELAND LTD. ACRYSOF IQ ASPHERIC IOL WITH THE ULTRASERT PRE-LOADED DELIVERY SYSTEM; INTRAOCULAR LENS Back to Search Results
Model Number AU00T0
Medical Device Problem Code Failure to Eject (4010)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been no other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A facility representative reported that during an intraocular lens (iol) implant procedure, the plunger was override the iol.No further information available.
 
Additional Manufacturer Narrative
The product was returned for analysis and the reported complaint was observed.Additional observations were as follows: the device is returned in the blister tray in the carton.The lens stop and the plunger stop has been removed.The correct nozzle confirmed on the device.Viscoelastic is observed in the device.The plunger and the lens are advanced into the mid nozzle and the plunger is advanced under the lens.The lens is misfolded.Additional information: the complainant indicates the use of non company viscoelastic as a viscoelastic, which is not qualified for use with associated company device.While we are unable to determine the origin of the reported complaint, our observations reasonably suggest that it is not manufacturing related.Based on the information provided by the customer, the most likely root cause is failure to follow ifu (instructions for use), as the customer states the use of non-qualified viscoelastic.The use of an unqualified ovd (ophthalmic viscosurgical device) may cause damage to the lens and potential complications during the implantation process.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ ASPHERIC IOL WITH THE ULTRASERT PRE-LOADED DELIVERY SYSTEM
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI  00000
Manufacturer (Section G)
ALCON LABORATORIES IRELAND LTD.
cork business&technology park
model farm road
cork 00000
EI   00000
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15113605
Report Number9612169-2022-00356
Device Sequence Number20199806
Product Code HQL
UDI-Device Identifier00380652394697
UDI-Public00380652394697
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2023
Device Model NumberAU00T0
Device Lot Number21282992
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/04/2022
Supplement Date Received by Manufacturer09/26/2022
Initial Report FDA Received Date07/27/2022
Supplement Report FDA Received Date10/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/09/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
RAYNER VISCOELASTIC OPHTEIS.
Patient SexUnknown
-
-