BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number H7493941838400
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/07/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that stent dislodgement occurred.A 4.00 x 38mm synergy xd drug-eluting stent was advanced to treat the lesion.However, during the procedure, the stent came off the delivery system.The device was removed from the patient body along with the stent delivery system.The procedure was completed with a new stent.No patient complications reported.
 
Event or Problem Description
It was reported that stent dislodgement occurred.A 4.00 x 38mm synergy xd drug-eluting stent was advanced to treat the lesion.However, during the procedure, the stent came off the delivery system.The device was removed from the patient body along with the stent delivery system.The procedure was completed with a new stent.No patient complications reported.It was further reported that the stent came off the balloon during removal of plastic protector and was never entered into the patient's body.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: synergy xd mr us 4.00 x 38mm stent delivery system (sds), catheter was returned for analysis.A visual examination of the stent found stent damage along most of the length of the stent profile with struts pulled distally from the fourth strut on the proximal end over the balloon cone and tip.The undamaged stent outer diameter was measured and the result was within max crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of distal tip damage.A visual and tactile examination of the hypotube shaft found multiple kinks.A visual and tactile examination of the outer and mid-shaft section and a visual examination of the inner lumen found no issues along the shaft polymer extrusion.No other issues were identified during analysis.
 
Event or Problem Description
It was reported that stent dislodgement occurred.A 4.00 x 38mm synergy xd drug-eluting stent was advanced to treat the lesion.However, during the procedure, the stent came off the delivery system.The device was removed from the patient body along with the stent delivery system.The procedure was completed with a new stent.No patient complications reported.It was further reported that the stent came off the balloon during removal of plastic protector and was never entered into the patient's body.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY XD
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15113847
Report Number2134265-2022-08153
Device Sequence Number5589524
Product Code NIQ
UDI-Device Identifier08714729981176
UDI-Public08714729981176
Combination Product (Y/N)Y
Initial Reporter StateAR
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/25/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/05/2024
Device Model NumberH7493941838400
Device Lot Number0028648857
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 07/19/2022
Supplement Date Received by Manufacturer07/26/2022
08/24/2022
Initial Report FDA Received Date07/27/2022
Supplement Report FDA Received Date08/05/2022
08/25/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/05/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-