Manufacturer's investigation conclusion: the reported event of damage to the black and white power cables of the system controller and the reported low voltage events were not confirmed.The heartmate 3 system controller (serial #: (b)(6)) was not returned for analysis and no log files were submitted for review.Multiple attempts were made to obtain additional information regarding the product return; however, no additional information was provided.If the product is returned in the future, the event will be updated, and product analysis will be performed at that time.The root causes of the reported events were unable to be determined in this analysis.The device history records were reviewed and the records revealed the heartmate 3 system controller was manufactured in accordance with manufacturing and qa specifications.Heartmate iii instructions for use section 7 ¿ ¿alarms and troubleshooting¿ and heartmate iii patient handbook section 5 ¿ ¿alarms and troubleshooting¿ explains how to properly interpret and troubleshoot all alarms, including low voltage advisory alarms.Heartmate iii instructions for use section 8 ¿ ¿equipment storage and care¿ and heartmate iii patient handbook section 6 ¿ ¿caring for equipment¿ explain how to properly care for the equipment.The patient handbook cautions the users to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.No further information was provided.The manufacturer is closing the file on this event.
|