| Model Number |
R SERIES |
| Medical Device Problem Code |
Unable to Obtain Readings (1516)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that while attempting to cardiovert a patient (age & gender unknown), the device displayed an "ecg disabled" message.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation; the customer's report was observed during review of the device's history log.The device was put through extensive testing which including defib cycling, impedance testing, bench handling, environmental chamber testing and sync testing through electrode pads and paddles without duplicating the report.An internal inspection resulted in no findings.The analog board was replaced as a precaution.The device passed the final test procedure, was recertified, and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
| |
|