| Model Number |
N/A |
| Medical Device Problem Codes |
Device Dislodged or Dislocated (2923); Naturally Worn (2988)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
07/08/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Medical product: catalog #: 110030776, cr 40mm glenosphere std cocr, lot # 65323728.Catalog #: 110030776, 40mm versa-dial glen std, lot # 65323728.Catalog #: 118000, 25mm versa-dial taper adaptor, lot # 909850.Catalog #: 110031400, hmrl tray std +5, lot # 65255146.Catalog #: 115395, comp rvs cntrl 6.5x25mm, lot # 182630.The customer has not indicated whether the product will be returned to zimmer biomet for investigation or not.Once this information is obtained a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001822565-2022-02230.
|
| |
|
Event or Problem Description
|
|
It was reported that patient had an initial surgery on an unknown date.Subsequently, about 2 weeks ago the patient was revised due to the glenosphere being dislocated.It was stated that there is also significant wear on the humeral poly.
|
| |
|
Event or Problem Description
|
|
It was further reported that the patient underwent a revision approximately 7 months post initial implantation.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Reported event was confirmed as visual examination of the returned products identified that the glenosphere, a taper adapter, a screw and a poly bearing were returned.There is damage noted on the glenosphere, bearing, and the tray.Device history record was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Search Alerts/Recalls
|