• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET 3I CERTAIN® GINGIHUE® POST 5MM(D) X 6MM(P) X 2MM(H); ABUTMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET 3I CERTAIN® GINGIHUE® POST 5MM(D) X 6MM(P) X 2MM(H); ABUTMENT Back to Search Results
Model Number IWPP562G
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/06/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the implant was removed due to peri-implantitis at the implant site.Customer reported implant had been placed in area #20.Abutment was inserted after 4 months of healing; after one week with provisional crown abute was noted to be loosed.Could not tighten abutment to implant to get platform to seat.
 
Additional Manufacturer Narrative
Zimmerbiomet complaint number (b)(4).
 
Event or Problem Description
No further event information is available at the time of this report.
 
Additional Manufacturer Narrative
Zimmer biomet complaint number (b)(4).The reported device was returned for investigation.Visual evaluation of the returned product identified no damage and engaged inside in-house implant.Functional testing was done using an in-hose implant and was able to engage.No pre-existing conditions were noted on the product experience report (per).X-ray or picture images were not provided.Complainant reported that after one week with provisional crown abutment was noted to be loosed.Could not tighten abutment to implant to get platform to seat.The reported complaint could not be verified.A definitive root cause for this complaint could not be determined.A device history record (dhr)review was performed and no related deviations or nonconformance's were noted.Also, a complaint history search was performed using our complaint handling system and there were no related complaints for this product lot.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CERTAIN® GINGIHUE® POST 5MM(D) X 6MM(P) X 2MM(H)
Common Device Name
ABUTMENT
Manufacturer (Section D)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
BIOMET 3I
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
susanne taylor
4555 riverside drive
palm beach gardens, FL 33410
5617766700
MDR Report Key15115078
Report Number0001038806-2022-01141
Device Sequence Number5849978
Product Code NHA
UDI-Device Identifier00844868009030
UDI-Public(01)00844868009030(17)270314(10)1252904
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
K072642
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date (Section B) 01/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberIWPP562G
Device Catalogue NumberIWPP562G
Device Lot Number1252904
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2022
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/12/2022
Supplement Date Received by Manufacturer12/27/2022
Initial Report FDA Received Date07/27/2022
Supplement Report FDA Received Date01/10/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age85 YR
Patient SexMale
-
-