• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION STERLING; CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION STERLING; CATHETER, PERCUTANEOUS Back to Search Results
Model Number 24700
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/21/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon rupture occurred.The 75% stenosed target lesion was located in moderately tortuous and non-calcified left forearm vein.A 7.0mmx40mmx80cm (4f) sterling balloon catheter was advanced and dilated the central side without problem.The same device was used to expand the lesion on the peripheral side.The balloon was inflated twice.However, during the second inflation at 6 atmospheres, the lesion could not be dilated and the elution of the contrast medium from the balloon was recognized as a ruptured.It was noted that there was no appearance of a rupture when the central side expands.The device was removed without any problem.The procedure was completed with another of same device.There were no patient complications nor injuries reported and the patient condition was good.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of a sterling balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual and microscopic examination revealed no damages.The balloon folds were loosely folded and there is contrast present inside the device.Functional testing was performed by inflating the device to rated burst pressure and it was able to hold pressure.Inspection of the remainder of the device presented no other damage or irregularities.
 
Event or Problem Description
It was reported that balloon rupture occurred.The 75% stenosed target lesion was located in moderately tortuous and non-calcified left forearm vein.A 7.0mmx40mmx80cm (4f) sterling balloon catheter was advanced and dilated the central side without problem.The same device was used to expand the lesion on the peripheral side.The balloon was inflated twice.However, during the second inflation at 6 atmospheres, the lesion could not be dilated and the elution of the contrast medium from the balloon was recognized as a ruptured.It was noted that there was no appearance of a rupture when the central side expands.The device was removed without any problem.The procedure was completed with another of same device.There were no patient complications nor injuries reported and the patient condition was good.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERLING
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15115347
Report Number2124215-2022-27306
Device Sequence Number14841306
Product Code DQY
UDI-Device Identifier08714729124108
UDI-Public08714729124108
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K053116
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2025
Device Model Number24700
Device Catalogue Number24700
Device Lot Number0028961170
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/21/2022
Supplement Date Received by Manufacturer08/18/2022
Initial Report FDA Received Date07/27/2022
Supplement Report FDA Received Date08/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDEWIRE USED: GLITTER 160; GUIDEWIRE USED: GLITTER 160; INTRODUCER SHEATH USED: MEDIKIT 4FR; INTRODUCER SHEATH USED: MEDIKIT 4FR
Patient SexMale
-
-